Water quality is essential for a wide range of applications, from industrial processes and drinking water treatment to environmental monitoring and research. One of the key indicators of water quality is the presence of organic carbon, which can significantly impact both human health and the ecosystem. In this blog post, Metash will share the working principles and methods of TOC total organic carbon analyzer for measuring water quality.
TOC Total Organic Carbon Analyzer Measurement Methods
TOC analyzers measure the concentration of organic carbon in water through several methods, with the most common ones being oxidation-based and non-oxidation-based techniques. Each method is employed based on the type of sample and the required sensitivity. Here, we focus on oxidation-based methods, which are the most widely used in the field.
1. High-Temperature Combustion Method (HTC)
The High-Temperature Combustion Method is one of the most reliable and precise methods for TOC measurement. In this method, the water sample is first acidified to remove inorganic carbon, primarily in the form of dissolved carbonates and bicarbonates. Once the sample is acidified, it is then passed through a combustion furnace at temperatures between 680°C and 900°C.
At this high temperature, the organic carbon present in the sample is oxidized into carbon dioxide (CO₂). The CO₂ produced is then measured by an infrared (IR) detector, which quantifies the amount of carbon dioxide released, providing a direct correlation to the organic carbon concentration in the sample.
The High-Temperature Combustion Method is highly effective for water samples with low to moderate levels of organic carbon and provides very accurate results. The high temperature ensures complete oxidation of even complex organic molecules, which is particularly beneficial for industrial applications requiring precise water quality monitoring.
2. Ultraviolet (UV) Oxidation Method
The UV oxidation method is a less aggressive technique for measuring TOC, often used when lower levels of organic carbon need to be quantified, or when a sample contains sensitive components that may be affected by high temperatures.
In this process, the sample is first acidified to remove inorganic carbon. Then, the sample is exposed to ultraviolet (UV) light in the presence of an oxidizing agent, such as ozone or persulfate. The UV radiation excites the organic molecules, facilitating the oxidation of organic carbon to CO₂. The CO₂ is then measured using a non-dispersive infrared (NDIR) detector, similar to the HTC method.
The UV oxidation method is beneficial in situations where high temperatures could degrade certain sensitive organic compounds or when there is a need for a faster analysis. However, this method may not be as effective for samples with high concentrations of organic carbon or for complex organic mixtures.
3. Persulfate Oxidation Method
In the persulfate oxidation method, the sample is combined with a persulfate reagent, which is a strong oxidizing agent. The persulfate oxidizes the organic carbon in the sample to CO₂, which is then measured using a detector. The persulfate oxidation method is often used for more challenging samples, including those with high turbidity or higher levels of organic carbon.
This method is particularly useful for environmental monitoring where sample diversity is high, and it can achieve low detection limits, making it suitable for trace analysis in natural water bodies.
Key Components of a TOC Analyzer
TOC analyzers consist of several key components that work together to measure the organic carbon content in water. These components include:
1. Sample Introduction System: The sample is introduced into the analyzer either by direct injection or through an automated sampling system. In some cases, pre-treatment of the sample, such as acidification to remove inorganic carbon, is performed.
2. Oxidation Chamber: This is where the organic carbon in the sample is oxidized to CO₂. Depending on the method used, this chamber may involve a high-temperature furnace, a UV light source, or a persulfate reagent.
3. Detector: After oxidation, the CO₂ produced is detected using an infrared (IR) detector or a non-dispersive infrared (NDIR) sensor. The detector measures the concentration of CO₂, which is directly proportional to the organic carbon content in the sample.
4. Data Processing Unit: The analyzer' s software processes the data collected from the detector and calculates the TOC concentration in the sample. The result is displayed in either milligrams per liter (mg/L) or parts per million (ppm), depending on the sensitivity of the instrument.
Conclusion
Total Organic Carbon analyzers are indispensable tools for measuring organic contaminants in water. By accurately quantifying the organic carbon content, TOC analyzers provide essential data for water quality assessment, ensuring that water used in industrial processes, drinking supplies, and environmental systems meets the necessary standards. With various methods available, from high-temperature combustion to UV oxidation, these analyzers offer flexibility and precision for a wide range of applications.